Preliminary Results of Quality Control for Linear Accelerator at Oncology Center in Nouakchott

El-Mouna, Ould and Moussa, Ould and Abdelmajid, Choukri and Jilali, Ghassoun and Keltoum, Hakam and Slimane, Semghouli (2017) Preliminary Results of Quality Control for Linear Accelerator at Oncology Center in Nouakchott. British Journal of Medicine and Medical Research, 22 (2). pp. 1-8. ISSN 22310614

[thumbnail of Mouna2222017BJMMR33918.pdf] Text
Mouna2222017BJMMR33918.pdf - Published Version

Download (199kB)

Abstract

The National Center of Oncology in Nouakchott is equipped with a linear accelerator which provides two energies in photon regime 6 MV and 18 MV.

The basic dosimetric data required for quality of operation, in the clinical routine of the accelerator for photon beams are percentage depth dose (PDD) curves and beam profile (BP) curves.

The aim of this work is to measure the percentage depth dose (PDD) and beam/dose profile (BP) for 6 MV and 18 MV photon beams for different field sizes and for different Skin-Source-Distance (SSD), and to compare the measured results with the results calculated by Treatment Planning Systems (TPS).

The measured results of (PDD) and (BF) are consistent with those calculated by TPS, which allows us to conclude that the linear accelerator at oncology Center in Nouakchott is in good condition.

Other Parameters: (energy, (SSD) and field sizes) have been also studied in this work, in order to perform a comprehensive quality study for this accelerator a unique oncology at Mauritania.

Item Type: Article
Subjects: Middle Asian Archive > Medical Science
Depositing User: Managing Editor
Date Deposited: 08 May 2023 07:10
Last Modified: 28 Aug 2024 14:16
URI: http://library.eprintglobalarchived.com/id/eprint/424

Actions (login required)

View Item
View Item